Accurate EM-TV Algorithm in PET with Low SNR
PET measurements of tracers with a lower dose rate
or short radioactive half life suffer from extremely low SNRs.
In these cases standard reconstruction methods (OSEM, EM,
filtered backprojection) deliver unsatisfactory and noisy results.
Here, we propose to introduce nonlinear variational methods into
the reconstruction process to make an efficient use of a-priori
information and to attain improved imaging results. We illustrate
our technique by evaluating cardiac H2
15O measurements. The
general approach can also be used for other specific goals
allowing to incorporate a-priori information about the solution
with Poisson distributed data.